Do I need a thermoregulator to heat and cool Mya 4?

No, you do not need a thermoregulator for your Mya 4.


The heating of Mya 4 heating is achieved using heating cartridges in all four zones. This provides fast and precise heating.


The active cooling is provided by Peltier technology and a cooled water supply is required for the heat exchanger.


Recommended Supply:

  • Coolant Supply Mains water or ethylene glycol mix (max 30%)
  • Max. coolant temperature 15 °C (up to 35 °C, however performance will be affected)
  • Min. coolant temperature 5 °C
  • Max. flow rate 5 L/min
  • Min. flow rate 2 L/min
  • Max. coolant pressure 2 bar
  • Min. coolant pressure 0.3 bar

  • If using a thermoregulator (both specs below required):

    Min. cooling power at 15 °C 1.5 kW

    Min. cooling power at 0 °C 1.0 kW